Biological Safety Cabinet TRBC-602

Biological Safety Cabinet TRBC-602

Biological Safety Cabinet TRBC-602 Is a closed working environment that can effectively protect against corrosion caused by corrosive substances and disinfectants. Digital display panels are used in all steel-framed workplaces. This includes UV lamps, fluorescent lights, and differential displays. There are HEPA filters and primary filters. HEPA filters contribute to a clean working environment. It is a sterilization method linked with safety functions. Discharges 100% of the airflow outside the cross-contamination.

In Stock

Average Wind Speed 0.38 ± 0.025 m/s, 0.26 ± 0.025m/s, 0.27 ± 0.025m/s
Front Suction Wind Speed 0.55m ± 0.025m/s (30% Efflux)
High-Efficiency Filter Specification And Quantity 1280×490×50×1, 820×380×70×1
Maximum Power Consumption 600 W
Noise ≤65 dB(A)
Number Of Colonies ≤0.5pcs/dish hour (Φ90mm culture plate)
Overall Dimension W×D×H 1500×800×2100 mm
Specification And Quantity Of Fluorescent Lamp/Ultraviolet Lamp 30W×2/30W×1
Vibration Half Peak ≤3 μm
Voltage AC single-phase, 220 V/50 Hz
Weight 250 Kg
Working Area W1×D1×H1 1340×650×620 mm
It is built with air curtain isolation in mindBy discharging 100% of the airflow outside, negative pressure vertical laminar flow, no need to install pipes
It prevents internal and external cross-contamination
The glass door can be moved up and down, and it can be positioned in any direction
It is simple to use and can be sterilized completely closed
W0ith the prompting of the positioning height, limit alarm
The work area's power output socket is equipped with a waterproof socket and a sewage interface for the operator's convenience
A special filter installed at the exhaustion of air to control pollution emissions
The working environment is made of high-quality stainless steel that is smooth, seamless, and corrosion-resistant
It can be disinfected quickly and thoroughly, and it can resist corrosion from corrosive agents and disinfectants
An LED LCD panel controls it
Built-in UV lamp protection device, which can only turn on when the safety door is shutdown
With a DOP detection port and a built-in differential pressure gauge, that tilts 10˚ by the human body design concept
Biological Safety Cabinet TRBC-601

Biological Safety Cabinet TRBC-601

  • Number Of Colonies : ≤0.5pcs/dish hour (Φ90mm culture plate)
  • Average Wind Speed : 0.38 ± 0.025 m/s, 0.26 ± 0.025m/s, 0.27 ± 0.025m/s
  • Front Suction Wind Speed : 0.55m ± 0.025m/s (30% Efflux)
Biological Safety Cabinet TRBC-603

Biological Safety Cabinet TRBC-603

  • Number Of Colonies : ≤0.5pcs/dish hour (Φ90mm culture plate)
  • Average Wind Speed : 0.38 ± 0.025 m/s, 0.26 ± 0.025m/s, 0.27 ± 0.025m/s
  • Front Suction Wind Speed : 0.55m ± 0.025m/s (30% Efflux)
Biological safety cabinet TRBC-604

Biological safety cabinet TRBC-604

  • The Average Wind Speed- Inner Side : 0.38 + -0.025m/s
  • The Average Wind Speed- Middle : 0.26 + -0.025 m/S
  • The Average Wind Speed-Inside : 0.35 + -0.025 m/s