Biological Safety Cabinet TRBC-602

Download

Overview

Biological Safety Cabinet TRBC-602 Is a closed working environment that can effectively protect against corrosion caused by corrosive substances and disinfectants. Digital display panels are used in all steel-framed workplaces. This includes UV lamps, fluorescent lights, and differential displays. There are HEPA filters and primary filters. HEPA filters contribute to a clean working environment. It is a sterilization method linked with safety functions. Discharges 100% of the airflow outside the cross-contamination.

Specifications :

Average Wind Speed 0.38 ± 0.025 m/s, 0.26 ± 0.025m/s, 0.27 ± 0.025m/s
Front Suction Wind Speed 0.55m ± 0.025m/s (30% Efflux)
High-Efficiency Filter Specification And Quantity 1280×490×50×1, 820×380×70×1
Maximum Power Consumption 600 W
Noise ≤65 dB(A)
Number Of Colonies ≤0.5pcs/dish hour (Φ90mm culture plate)
Overall Dimension W×D×H 1500×800×2100 mm
Specification And Quantity Of Fluorescent Lamp/Ultraviolet Lamp 30W×2/30W×1
Vibration Half Peak ≤3 μm
Voltage AC single-phase, 220 V/50 Hz
Weight 250 Kg
Working Area W1×D1×H1 1340×650×620 mm
Features :
  • It is built with air curtain isolation in mindBy discharging 100% of the airflow outside, negative pressure vertical laminar flow, no need to install pipes
  • It prevents internal and external cross-contamination
  • The glass door can be moved up and down, and it can be positioned in any direction
  • It is simple to use and can be sterilized completely closed
  • W0ith the prompting of the positioning height, limit alarm
  • The work area's power output socket is equipped with a waterproof socket and a sewage interface for the operator's convenience
  • A special filter installed at the exhaustion of air to control pollution emissions
  • The working environment is made of high-quality stainless steel that is smooth, seamless, and corrosion-resistant
  • It can be disinfected quickly and thoroughly, and it can resist corrosion from corrosive agents and disinfectants
  • An LED LCD panel controls it
  • Built-in UV lamp protection device, which can only turn on when the safety door is shutdown
  • With a DOP detection port and a built-in differential pressure gauge, that tilts 10˚ by the human body design concept